- 28 -
Secondary display parameter is
ș.
Press “ D/Q/ESR/ș ” key
Back to parameter D
measurement
Ɣ D/Q/ESR/ș setting is not applied to resistance
measurement- Auto-R mode and DCR mode.
In Auto-R mode, no parameter
measurement on the secondary
display.
In DCR mode, no parameter
measurement on the secondary
display.
- 29 -
10. SER/PAL
y In LCR Auto mode, this key has no function.
y Press the “SER/PAL” key to select parallel or series mode,.
y When any one of primary parameter measurement function
is selected, parallel or series mode will be selected
automatically based on the total equivalent impedance
measured.
y If the impedance is more than 100kȍ, will be in the parallel
mode automatically. Lp, Cp or Rp appears on the display.
y If the impedance is less than 100kȍ, will be in the series
mode automatically. Ls, Cs or Rs appears on the display.
Example : In Cs and Cp testing mode
When measure capacitance in
series mode.
“Cs” will appear.
Press “SER/PAL ” key
When measure capacitance In
Parallel mode.
“Cp” is shown.
Press “SER/PAL ” key
Back to Series mode.
“Cs” is shown.
y
The annunciators for Ls/Lp/Cs/Cp/Rs/Rp will be indicated based
on LCR measurement mode.